What a PMS Plan Must Contain
Annex III of the MDR sets out the minimum content of a PMS Plan. Put simply, it comes down to four elements:
- Which sources are used – Incident Reportsuser feedback, scientific literature, registers, and publicly available information on comparable products
- How this information is assessed through indicators and thresholds that feed into the ongoing benefit-risk assessment under Annex I, Section 3
- When a trend becomes reportable that is, a statistically significant increase in the frequency or severity of non-serious incidents and expected undesirable side-effects under Article 88 MDR
- How the manufacturer responds in a serious case which procedures are used to inform authorities and Notified bodies and to initiate corrective action
A Class III pacemaker needs different surveillance methods to a Class I compression stocking from the German Medical Technical Aids-Register. The MDR does not prescribe a fixed scheme for this, but it does require that the choice of method and interval be traceable and justified – and in practice, this justification is surprisingly often missing.
PMS and PMCF – Two Terms that are Easily Confused
PMS and PMCF are frequently mixed up, though the distinction is straightforward. PMS is the overarching framework, gathering all experience with the product: complaints, authority data, information on comparable products. PMCF is one part of that framework, looking specifically at clinical data: does the evidence continue to confirm the safety and performance of the product, or are new risks emerging? Both processes build on the clinical evaluation carried out before CE marking , which already indicates what needs particular attention after market launch. What a PMCF plan must specifically contain is set out in Annex XIV, Part B of the MDR.
When a PMS Report is Enough – and when a PSUR is Required
Not every product needs the same type of report. For Class I devices, a PMS report under Article 85 MDR is sufficient, updated as needed. From Class IIa upwards, a Periodic Safety Update Report (PSUR) becomes mandatory: at least every two years for Class IIa, and at least annually for Class IIb and III. Manufacturers who overlook this distinction risk, in the worst case, a finding at their next Auditbecause a PSUR that was due simply does not exist.
Where the Data Comes from – and Where it Usually Gets Stuck
Once a product is on the market, active data collection begins. Alongside incident reports and customer feedback, this includes trend reports, scientific literature, relevant registers, and publicly available information on comparable products. In practice, the last two points are the biggest time sinks: literature databases and registers rarely yield to a quick keyword search, and judging which information on similar products is genuinely relevant takes experience.
This is precisely where BEO BERLIN comes in. We research the relevant literature databases and registers, assess the findings from a regulatory perspective, and prepare them so they slot directly into the PMS report or PSUR. Literature research then becomes a plannable building block, rather than a last-minute scramble before the reporting deadline.
From Data Collection to Documentation
A PMS system only demonstrates its value once something actually follows from the data collected: necessary preventive and corrective actions have to be identified, implemented, and checked for effectiveness. The results feed into the PMS report or PSUR, depending on risk class. Neither is a box to tick at year end; both are the visible outcome of a process that, ideally, was set up carefully long before the reporting date.
Where Does your PMS Plan Stand?
If you are not entirely sure where your product stands in the PMS process, you are not alone. For legacy products with documentation that has grown over time, this is the rule rather than the exception. We are happy to look with you at whether your PMS plan meets the requirements of Annex III, which type of report applies to your product, and where PMS and PMCF should be more clearly separated. Get in touchif you would value a second, expert opinion on this.
(1 Rating(s), average: 5.00 out of 5)