Technical Documentation under the MDR: Structure, Content and Common Weaknesses

The MDR requires the documentation to be presented in a clear, organised, readily searchable and unambiguous manner. There is more to this wording than a formality.

The Technical Documentation is the evidence that a medical device meets the requirements of the Medical Devices Regulation (MDR) . Under Article Art. 10 MDR , every manufacturer must draw it up and keep it up to date, regardless of the risk class. Even for Class I devices, where no notified body is involved, it must be in place before the manufacturer issues the EU-Declaration of Conformity . Two annexes define its content and structure: Annex II for the documentation on the device itself, and Annex III for post-market surveillance.

A Chain of Reasoning, Not a Filing Cabinet

Anyone reviewing technical documentation does not read it chapter by chapter as a collection of individual documents. What is assessed is whether the parts fit together. The intended purpose defines who the device is intended for and what it is intended to do. The classification, the applicable safety and performance requirements, the risks, the tests and the clinical data all follow from it. Each of these building blocks must refer to the others.

In practice, problems therefore rarely arise because a document is missing entirely. More often, the information does not match: the instructions for use state an indication that is not supported by the clinical evaluation, or the risk management file lists measures whose effectiveness no test demonstrates. Anyone setting up or revising documentation should therefore start with a clearly defined intended purpose and classification.

What Annex II Requires

Annex II divides the technical documentation into six areas:

  • Device description and specification: intended purpose, users, patient population, indications and contraindications, principles of operation, classification with justification, variants, accessories, materials in contact with the body and technical specifications. This includes the Basic-UDI-DIor, if the device is not yet identified through the UDI system, another clear means of identification. The MDR also requires an overview of previous generations of the device and similar devices on the market.
  • Information to be supplied by the manufacturer: Label, packaging and Instructions for Use in the languages of the Member States where the device is to be sold.
  • Design and manufacturing information: the design stages, the manufacturing processes and their validation, ongoing monitoring and final testing, and all sites where design and manufacturing take place, including Supplier and subcontractors.
  • General safety and performance requirements: for each requirement in Annex I, whether it applies, which method is used to demonstrate conformity, which harmonised standards or common specifications are applied and where the evidence can be found in the documentation. Requirements that do not apply must be justified.
  • Benefit-risk analysis and risk management: the analysis itself, the solutions adopted and the results of risk management.
  • Product verification and validation: the results and critical analyses of all tests and studies. These include pre-clinical data, for example on Biocompatibilitysoftware, stability and performance, as well as the Clinical Evaluation Report and report and the post-market clinical follow-up (PMCF) plan, or a justification of why it is not required.

For certain devices, such as sterile devices, devices with a measuring function or devices incorporating a medicinal substance, additional information is required.

The Often Overlooked Part: Annex III

The technical documentation also includes the documentation on post-market surveillance. Annex III requires a Post-Market Surveillanceplan describing how the manufacturer collects and evaluates data on Incident Reporting, specialist literature and similar devices. Depending on the risk class, a report is also required: manufacturers of Class I devices draw up a post-market surveillance report and update it when necessary. From Class IIa upwards, a periodic safety update report (PSUR) is required, updated at least every two years for Class IIa and at least annually for Classes IIb and III.

This part makes it clear that the technical documentation is never finished. Findings from the market feed back into risk management, the clinical evaluation and the instructions for use.

Keeping It Up to Date, Not Just Drawing It Up

The MDR obliges manufacturers to keep the technical documentation up to date on an ongoing basis. Triggers include changes to the device, new data from post-market surveillance, revised standards and new guidance. In addition, the documentation must remain available to the authorities for at least ten years after the last device has been placed on the market, and for at least 15 years for implantable devices.

With devices that have been on the market for a long time in particular, documentation easily becomes outdated without anyone noticing: standards are withdrawn, references no longer match, and changes to the device have only been partly carried through. This often only comes to light during an Audit .

Review It Before the Regional Authority or the Notified Body does

BEO BERLIN supports manufacturers in two ways: with the structured set-up of new technical documentation and with the review of existing documentation. In a review, we check the documentation against the requirements of Annexes II and III, paying particular attention to whether the building blocks are consistent with each other. The result is a clear list of what is missing, what is contradictory and what can be closed with reasonable effort. On request, we then support you through the conformity assessment.

If a recertification, an audit or the launch of a new device is coming up, this review is worth doing beforehand. Have BEO BERLIN review your technical documentation for completeness and consistency.